An '''electron microscope''' is a type of [[microscope]] that uses [[electron]]s as a way to illuminate and create an image of a specimen. It has much higher [[magnification]] and [[resolving power]] than a [[Optical microscope|light microscope]], with magnifications up to about two million times, compared to about two thousand that can be achieved with visible light.
+
[[Image:Microscope-IMG 0518.jpg|thumb|300px|An electron microscope.]]
  
This high magnification allows electron microscopes to see smaller objects and greater detail in these objects.  Unlike a light microscope, which uses glass lenses to focus light, the electron microscope uses electrostatic and electromagnetic lenses to control the illumination and imaging of the specimen.
+
An '''electron microscope''' is a type of [[microscope]] that uses [[electron]]s to illuminate a specimen and create an enlarged image. Electron microscopes have much greater [[Angular resolution|resolving power]] than [[Optical microscope|light microscopes]] and can obtain much higher [[magnification]]s. Some electron microscopes can magnify specimens up to 2 million times, while the best light microscopes are limited to magnifications of 2000 times. Both electron and light microscopes have resolution limitations, imposed by their wavelength. The greater resolution and magnification of the electron microscope is due to the wavelength of an electron, its [[de Broglie wavelength]], being much smaller than that of a [[Light|light photon]], [[electromagnetic radiation]].
 +
{{toc}}
 +
The electron microscope uses [[Electrostatic lens|electrostatic]] and [[electromagnetic]] lenses in forming the image by controlling the electron beam to focus it at a specific plane relative to the specimen in a manner similar to how a light microscope uses glass lenses to focus light on or through a specimen to form an image.
  
==History==  
+
==History==
[[Image:Elektronenmikroskop.jpg|frame|right|thumb|200px|An old model transmission electron microscope.]]
+
[[Image:Ernst Ruska Electron Microscope - Deutsches Museum - Munich-edit.jpg|thumb|right|Electron microscope constructed by [[Ernst Ruska]] in 1933.]]
  
[[Image:Ant_SEM.jpg|thumb|right|An image of an [[ant]] from a scanning electron microscope]]
+
{{readout|The first electron microscope prototype was built in 1931 by German engineers [[Ernst Ruska]] and [[Max Knoll]]|left}}.<ref name="RuskaNobel">Ernst Ruska, [http://nobelprize.org/nobel_prizes/physics/laureates/1986/ruska-autobio.html Ernst Ruska Nobel Prize autobiography,] Nobel Prize. Retrieved October 5, 2008.</ref> Although this initial instrument was only capable of magnifying objects by four hundred times, it demonstrated the principles of an electron microscope. Two years later, Ruska constructed an electron microscope that exceeded the resolution possible using an optical microscope.<ref name="RuskaNobel"/>
The first electron microscope prototype was built in [[1931]] by the German engineers [[Ernst Ruska]] and [[Max Knoll]].<ref>[http://nobelprize.org/nobel_prizes/physics/laureates/1986/ruska-autobio.html Ernst Ruska Nobel Prize autobiography]</ref> It was based on the ideas and discoveries of French physicist [[Louis de Broglie]]. Although it was primitive and not fit for practical use, the instrument was still capable of magnifying objects by four hundred times.
 
  
[[Reinhold Rudenberg]], the research director of [[Siemens AG|Siemens]], had patented the electron microscope in 1931, although Siemens was doing no research on electron microscopes at that time. In 1937 Siemens began funding Ruska and [[Bodo von Borries]] to develop an electron microscope. Siemens also employed Ruska's brother [[Helmut Ruska|Helmut]] to work on applications, particularly with biological specimens.<ref>{{cite web |url=http://nobelprize.org/nobel_prizes/physics/laureates/1986/ruska-autobio.html |title=Ernst Ruska Autobiography |accessdate=2007-02-06 |author=Ernst Ruska |year=1986 |publisher=Nobel Foundation |language=English }}</ref><ref>{{cite journal
+
[[Reinhold Rudenberg]], the research director of [[Siemens AG|Siemens]], had patented the electron microscope in 1931, although Siemens was doing no research on electron microscopes at that time. In 1937, Siemens began funding Ruska and [[Bodo von Borries]] to develop an electron microscope. Siemens also employed Ruska's brother [[Helmut Ruska|Helmut]] to work on applications, particularly with biological specimens.<ref name="RuskaNobel"/><ref>D.H. Kruger, P. Schneck, and H.R. Gelderblom, Helmut Ruska and the visualization of viruses, ''The Lancet'' 355 (9216): 1713–1717.</ref>
| author = DH Kruger, P Schneck and HR Gelderblom
 
| date = 13
 
| year = 2000
 
| month = May
 
| title = Helmut Ruska and the visualisation of viruses
 
| journal = The Lancet
 
| volume = 355
 
| issue = 9216
 
| pages = 1713-1717
 
| doi = 10.1016/S0140-6736(00)02250-9
 
| id =
 
| url = http://dx.doi.org/10.1016/S0140-6736(00)02250-9
 
| language = English}}
 
</ref>
 
  
In the same decade of 1930s [[Manfred von Ardenne]] pioneered the [[scanning electron microscope]] and his universal electron microscope.<ref>{{cite journal
+
In the same decade [[Manfred von Ardenne]] pioneered the [[scanning electron microscope]] and his universal electron microscope.<ref>M. von Ardenne and D. Beischer, Untersuchung von metalloxud-rauchen mit dem universal-elektronenmikroskop, ''Zeitschrift Electrochemie'' 46: 270–277.</ref>
| author = M von Ardenne and D Beischer
 
| date =
 
| year = 1940
 
| month =
 
| title = Untersuchung von metalloxud-rauchen mit dem universal-elektronenmikroskop  
 
| journal = Zeitschrift Electrochemie
 
| volume = 46
 
| issue =
 
| pages = 270-277
 
| language = German}}
 
</ref>
 
  
Siemens produced the first commercial TEM in 1939, but the first practical electron microscope had been built at the [[University of Toronto]] in [[1938]], by [[Eli Franklin Burton]] and students Cecil Hall, [[James Hillier]] and Albert Prebus.<ref>[http://web.mit.edu/Invent/iow/hillier.html MIT biography of Hillier]</ref>
+
Siemens produced the first commercial TEM in 1939, but the first practical electron microscope had been built at the [[University of Toronto]] in 1938, by [[Eli Franklin Burton]] and students Cecil Hall, [[James Hillier]], and Albert Prebus.<ref>Lemelson-MIT Program, MIT biography of Hillier.</ref>
  
Although modern electron microscopes can magnify objects up to two million times, they are still based upon Ruska's [[prototype]]. The electron microscope is an integral part of many laboratories. Researchers use it to examine biological materials (such as [[microorganisms]] and [[cell (biology)|cells]]), a variety of large [[molecules]], medical [[biopsy]] samples, [[metals]] and [[crystalline]] structures, and the characteristics of various surfaces. The electron microscope is also used extensively for inspection, quality assurance and failure analysis applications in industry, including, in particular, [[semiconductor device fabrication]].
+
Although modern electron microscopes can magnify objects up to two million times, they are still based upon Ruska's [[prototype]]. The electron microscope is an integral part of many laboratories. Researchers use them to examine biological materials (such as [[microorganisms]] and [[cell (biology)|cells]]), a variety of large [[molecules]], medical [[biopsy]] samples, [[metals]] and [[crystalline]] structures, and the characteristics of various surfaces. The electron microscope is also used extensively for inspection, quality assurance and failure analysis applications in industry, including, in particular, [[semiconductor device fabrication]].
  
== Electron microscope manufacturers ==
+
==Types==
 +
====Transmission Electron Microscope (TEM)====
 +
[[Image:Electron Microscope.png|right|upright|thumb|250px|Diagram of a transmission electron microscope.]]
  
Major manufacturers include:
+
The original form of electron microscopy, [[Transmission electron microscopy]] (TEM) involves a high [[voltage]] [[electron beam]] emitted by an [[electron gun]], usually fitted with a [[tungsten]] filament [[cathode]] as the electron source. The electron beam is accelerated by an [[anode]] typically at +100k[[eV]] (40 to 400 keV) with respect to the cathode, focused by [[electrostatic]] and [[electromagnetic]] lenses, and transmitted through a specimen that is in part transparent to electrons and in part [[Electron scattering|scatters]] them out of the beam. When it emerges from the specimen, the electron beam carries information about the structure of the specimen that is magnified by the [[objective lens]] system of the microscope. The spatial variation in this information (the "image") is recorded by projecting the magnified electron image onto a fluorescent viewing screen coated with a [[phosphor]] or [[scintillator]] material such as zinc sulfide. The image can be photographically recorded by exposing a [[photographic film]] or [[photographic plate|plate]] directly to the electron beam, or a high-resolution phosphor may be coupled by means of a [[fiber optic]] light-guide to the sensor of a CCD ([[charge-coupled device]]) camera. The image detected by the CCD may be displayed on a monitor or computer.
* [[Delong Group]]
 
* [[FEI Company]] – USA (merged with [[Philips]] Electron Optics)
 
* [[FOCUS GmbH]] – Germany
 
* [[Hitachi, Ltd.|Hitachi]] – Japan
 
* [[JEOL|JEOL Ltd.]] – Japan (Japan Electro Optics Laboratory)
 
* [[TESCAN]] – EU
 
* [[Zeiss|Carl Zeiss NTS GmbH]]
 
  
==Types ==
+
Resolution of the TEM is limited primarily by [[spherical aberration]], but a new generation of aberration correctors have been able to partially overcome spherical aberration to increase resolution. Software correction of spherical aberration for the High Resolution TEM [[HRTEM]] has allowed the production of images with sufficient resolution to show carbon atoms in diamond separated by only 0.89 [[ångström]] (89 [[picometer]]s) and atoms in silicon at 0.78 ångström (78 picometers)<ref>Paul Preuss, OÅM: World-Record Resolution at 0.78 Å, Berkeley Lab Currents.</ref><ref>P.D. Nellist, M.F. Chisholm, N. Dellby, O.L. Krivanek, M.F. Murfitt, Z.S. Szilagyi, A.R. Lupini, A. Borisevich, W.H. Sides, Jr., and S.J. Pennycook, [http://www.sciencemag.org/cgi/content/abstract/305/5691/1741 Direct Sub-Angstrom Imaging of a Crystal Lattice,] ''Science'' 305 (5691): 1741. Retrieved October 5, 2008.</ref> at magnifications of 50 million times.<ref>Department of Energy, [http://www.sc.doe.gov/bes/scale_of_things.html The Scale of Things,] Office of Basic Energy Sciences (BES). Retrieved October 5, 2008.</ref> The ability to determine the positions of atoms within materials has made the HRTEM an important tool for nano-technologies research and development.
 
 
====Transmission Electron Microscope (TEM) ====
 
{{main|Transmission electron microscopy}}
 
The original form of electron microscopy, [[Transmission electron microscopy]] (TEM) involves a high [[voltage]] electron beam emitted by a [[cathode]], usually a tungsten filament and focused by electrostatic and [[electromagnetic]] lenses. The electron beam that has been transmitted through a specimen that is in part transparent to electrons carries information about the inner structure of the specimen in the electron beam that reaches the imaging system of the microscope. The spatial variation in this information (the "image") is then magnified by a series of electromagnetic lenses until it is recorded by hitting a fluorescent screen, photographic plate, or light sensitive sensor such as a CCD ([[charge-coupled device]]) camera. The image detected by the CCD may be displayed in real time on a monitor or computer.
 
 
 
Resolution of the TEM is limited primarily by [[spherical aberration]], but a new generation of aberration correctors have been able to partially overcome spherical aberration to increase resolution. Software correction of spherical aberration for the High Resolution TEM [[HRTEM]] has allowed the production of images with sufficient resolution to show carbon atoms in diamond separated by only 0.89 [[ångström]] (89 [[picometer]]s) and atoms in silicon at 0.78 ångström (78 picometers)<ref>[http://www.lbl.gov/Publications/Currents/Archive/May-18-2001.html#_Hlk514817949 OÅM: World-Record Resolution at 0.78 Å], (May 18, 2001) Berkeley Lab Currents.</ref><ref>{{cite journal
 
| author = P. D. Nellist, M. F. Chisholm, N. Dellby, O. L. Krivanek, M. F. Murfitt, Z. S. Szilagyi, A. R. Lupini, A. Borisevich, W. H. Sides, Jr., S. J. Pennycook  
 
| date = 17
 
| year = 2004
 
| month = September
 
| title = Direct Sub-Angstrom Imaging of a Crystal Lattice
 
| journal = Science
 
| volume = 305
 
| issue = 5691
 
| pages = 1741
 
| doi = 10.1126/science.1100965
 
| id =
 
| url = http://www.sciencemag.org/cgi/content/abstract/305/5691/1741
 
| language = English}}
 
</ref> at magnifications of 50 million times.<ref>[http://www.sc.doe.gov/bes/scale_of_things.html The Scale of Things], DOE Office of Basic Energy Sciences (BES).</ref> The ability to determine the positions of atoms within materials has made the HRTEM an important tool for nano-technologies research and development.
 
  
 
====Scanning Electron Microscope (SEM) ====
 
====Scanning Electron Microscope (SEM) ====
{{Main|Scanning Electron Microscope}}
+
Unlike the TEM, where electrons of the high voltage beam form the image of the specimen, the [[Scanning Electron Microscope]] (SEM)<ref>D. McMullan, [http://www-g.eng.cam.ac.uk/125/achievements/mcmullan/mcm.htm Scanning Electron Microscopy, 1928-1965,] Microscopy Society of America. Retrieved October 5, 2008.</ref> produces images by detecting low energy secondary electrons which are emitted from the surface of the specimen due to excitation by the primary electron beam. In the SEM, the electron beam is rastered across the sample, with detectors building up an image by mapping the detected signals with beam position.
Unlike the TEM, where electrons of the high voltage beam form the image of the specimen, the [[Scanning Electron Microscope]] (SEM)<ref>[http://www-g.eng.cam.ac.uk/125/achievements/mcmullan/mcm.htm SCANNING ELECTRON MICROSCOPY 1928 - 1965]</ref> produces images by detecting low energy secondary electrons which are emitted from the surface of the specimen due to excitation by the primary electron beam. In the SEM, the electron beam is rastered across the sample, with detectors building up an image by mapping the detected signals with beam position.
+
[[Image:Ant SEM.jpg|thumb|right|An image of an [[ant]] from a scanning electron microscope]]
  
 
Generally, the TEM resolution is about an order of magnitude greater than the SEM resolution, however, because the SEM image relies on surface processes rather than transmission it is able to image bulk samples and has a much greater depth of view, and so can produce images that are a good representation of the 3D structure of the sample.
 
Generally, the TEM resolution is about an order of magnitude greater than the SEM resolution, however, because the SEM image relies on surface processes rather than transmission it is able to image bulk samples and has a much greater depth of view, and so can produce images that are a good representation of the 3D structure of the sample.
  
 
====Reflection Electron Microscope (REM) ====
 
====Reflection Electron Microscope (REM) ====
In addition there is a '''Reflection Electron Microscope''' (REM). Like TEM, this technique involves electron beams incident on a surface, but instead of using the transmission (TEM) or secondary electrons (SEM), the reflected beam is detected. This technique is typically coupled with [[RHEED|Reflection High Energy Electron Diffraction]] and ''Reflection high-energy loss spectrum (RHELS)''. Another variation is Spin-Polarized Low-Energy Electron Microscopy (SPLEEM), which is used for looking at the microstructure of [[magnetic domain]]s.<ref>http://ncem.lbl.gov/frames/spleem.htm</ref>
+
In the '''Reflection Electron Microscope''' (REM) as in the TEM, an electron beam is incident on a surface, but instead of using the transmission (TEM) or secondary electrons (SEM), the reflected beam of [[Elastic scattering|elastically scattered electrons]] is detected. This technique is typically coupled with [[RHEED|Reflection High Energy Electron Diffraction]] and ''Reflection high-energy loss spectrum (RHELS)''. Another variation is Spin-Polarized Low-Energy Electron Microscopy (SPLEEM), which is used for looking at the microstructure of [[magnetic domain]]s.<ref>National Center for Electron Microscopy, [http://ncem.lbl.gov/frames/spleem.html SPLEEM.] Retrieved October 5, 2008.</ref>
  
 
====Scanning Transmission Electron Microscope (STEM) ====
 
====Scanning Transmission Electron Microscope (STEM) ====
{{Main|Scanning Transmission Electron Microscope}}
+
The STEM rasters a focused incident probe across a specimen that (as with the TEM) has been thinned to facilitate detection of electrons scattered ''through'' the specimen. The high resolution of the TEM is thus possible in STEM. The focusing action (and aberrations) occur before the electrons hit the specimen in the STEM, but afterward in the TEM. The STEM's use of SEM-like beam rastering simplifies [[annular dark-field imaging]], and other analytical techniques, but also means that image data is acquired in serial rather than in parallel fashion.
  
The STEM combines the high resolution of the TEM with the beam rastering functions of the SEM, allowing a range of analytical techniques to be used that are not possible with conventional TEM.
+
== Sample preparation ==
 +
[[Image:Golden insect 01 Pengo.jpg|thumb|An insect coated in gold for viewing with a scanning electron microscope.]]
  
== Sample Preparation ==
 
[[Image:Golden insect 01 Pengo.jpg|thumb|An insect coated in gold, having been prepared for viewing with a scanning electron microscope.]]
 
 
Materials to be viewed under an electron microscope may require processing to produce a suitable sample. The technique required varies depending on the specimen and the analysis required:
 
Materials to be viewed under an electron microscope may require processing to produce a suitable sample. The technique required varies depending on the specimen and the analysis required:
*''Cryofixation'' – freezing a specimen so rapidly, to liquid nitrogen or even liquid helium temperatures, that the water forms [[Amorphous ice|vitreous (non-crystalline) ice]]. This preserves the specimen in a snapshot of its solution state. An entire field called [[cryo-electron microscopy]] has branched from this technique. With the development of [[cryo-electron microscopy|cryo-electron microscopy of vitreous sections]] (CEMOVIS), it is now possible to observe virtually any biological specimen close to its native state.
+
* Chemical [[fixation (histology)|Fixation]] for biological specimens aims to stabilize the specimen's mobile macromolecular structure by chemical crosslinking of [[protein]]s with [[aldehyde]]s such as [[formaldehyde]] and [[glutaraldehyde]], and [[lipid]]s with [[osmium tetroxide]].
*''Dehydration'' – replacing [[water]] with organic solvents such as [[ethanol]] or [[acetone]].  
+
* ''[[Cryofixation]]''—freezing a specimen so rapidly, to [[liquid nitrogen]] or even [[liquid helium]] temperatures, that the water forms [[Amorphous ice|vitreous (non-crystalline) ice]]. This preserves the specimen in a snapshot of its solution state. An entire field called [[cryo-electron microscopy]] has branched from this technique. With the development of cryo-electron microscopy of vitreous sections (CEMOVIS), it is now possible to observe virtually any biological specimen close to its native state.
*''Embedding'' – infiltration of the tissue with a [[resin]] such as [[araldite]] or [[epoxy]] for sectioning. After this embedding process begins, the specimen must be polished to a mirror-like finish using ultra-fine abrasives. The polishing process must be done accordingly, or it may lead to scratches imposing on the image quality.
+
* ''Dehydration''—[[freeze drying]], or replacement of [[water]] with organic solvents such as [[ethanol]] or [[acetone]], followed by [[critical point drying]] or infiltration with embedding resins.  
*''Sectioning'' – produces thin slices of specimen, semitransparent to electrons. These can be cut on an [[ultramicrotome]] with a [[diamond]] knife to produce very thin slices. [[Glass knives]] are also used because they can be made in the lab and are much cheaper.  
+
* ''Embedding, biological specimens''—infiltration of the tissue with a [[resin]] such as [[Araldite]] [[epoxy]] or [[acrylic]] resin followed by ultra-thin sectioning and staining.  
*''Staining'' – uses heavy metals such as [[lead]], [[uranium]] or [[tungsten]] to scatter imaging electrons and thus give contrast between different structures, since many (especially biological) materials are nearly "transparent" to electrons (weak phase objects). In biology, specimens are usually stained "en bloc" before embedding and also later stained directly after sectioning by brief exposure to aqueous (or alcoholic) solutions of the heavy metal stains.
+
* ''Embedding, materials''—After embedding in resin, the specimen is usually ground and polished to a mirror-like finish using ultra-fine abrasives. The polishing process must be performed carefully to minimize scratches and other polishing artifacts that reduce image quality.  
*''Freeze-fracture or freeze-etch'' – a preparation method particularly useful for examining lipid membranes and their incorporated proteins in "face on" view. The fresh tissue or cell suspension is frozen rapidly (cryofixed), then fractured by simply breaking or by using a microtome while maintained at liquid nitrogen temperature. The cold fractured surface (sometimes "etched" by increasing the temperature to about -100°C for several minutes to let some ice sublime) is then shadowed with evaporated platinum or gold at an average angle of 45° in a high vacuum evaporator. A second coat of carbon, evaporated perpendicular to the average surface plane is often performed to improve stability of the replica coating. The specimen is returned to room temperature and pressure, then the extremely fragile "pre-shadowed" metal replica of the fracture surface is released from the underlying biological material by careful chemical digestion with acids, hypochlorite solution or SDS detergent. The still-floating replica is thoroughly washed from residual chemicals, carefully fished up on EM grids, dried then viewed in the TEM.  
+
* ''Sectioning''—produces thin slices of specimen, semitransparent to electrons. These can be cut on an [[ultramicrotome]] with a [[diamond]] knife to produce ultrathin slices about 90 nm thick. [[Glass knives]] are also used because they can be made in the lab and are much cheaper.  
*''Ion Beam Milling'' – thins samples until they are transparent to electrons by firing [[ions]] (typically [[argon]]) at the surface from an angle and sputtering material from the surface. A subclass of this is [[Focused ion beam]] milling, where [[gallium]] ions are used to produce an electron transparent membrane in a specific region of the sample, for example through a device within a microprocessor. Ion beam milling may also be used for cross-section polishing prior to SEM analysis of materials that are difficult to prepare using mechanical polishing.  
+
* ''Staining''—uses heavy metals such as [[lead]], [[uranium]] or [[tungsten]] to scatter imaging electrons and thus give contrast between different structures, since many (especially biological) materials are nearly "transparent" to electrons (weak phase objects). In biology, specimens are usually stained "en bloc" before embedding and also later stained directly after sectioning by brief exposure to aqueous (or alcoholic) solutions of the heavy metal stains.
*''Conductive Coating'' – An ultrathin coating of electrically-conducting material, deposited either by high vacuum evaporation or by low vacuum sputter coating of the sample. This is done to prevent the accumulation of static electric fields at the specimen due to the electron irradiation required during imaging. Such coatings include gold, gold/palladium, platinum, tungsten, graphite etc. and are especially important for the study of specimens with the scanning electron microscope. Another reason for coating, even when there is more than enough conductivity, is to improve contrast, a situation more common with the operation of a FESEM (field emission SEM). When an osmium coater is used, a layer far thinner than would be possible with any of the previously mentioned sputtered coatings is possible.[http://www.2spi.com/catalog/osmi-coat.html]
+
* ''Freeze-fracture or freeze-etch''—a preparation method particularly useful for examining lipid membranes and their incorporated proteins in "face on" view. The fresh tissue or cell suspension is frozen rapidly (cryofixed), then fractured by simply breaking or by using a microtome while maintained at liquid nitrogen temperature. The cold fractured surface (sometimes "etched" by increasing the temperature to about -100°C for several minutes to let some ice sublime) is then shadowed with evaporated platinum or gold at an average angle of 45° in a high vacuum evaporator. A second coat of carbon, evaporated perpendicular to the average surface plane is often performed to improve stability of the replica coating. The specimen is returned to room temperature and pressure, then the extremely fragile "pre-shadowed" metal replica of the fracture surface is released from the underlying biological material by careful chemical digestion with acids, hypochlorite solution or SDS detergent. The still-floating replica is thoroughly washed from residual chemicals, carefully fished up on EM grids, dried then viewed in the TEM.  
 +
* ''Ion Beam Milling''—thins samples until they are transparent to electrons by firing [[ion]]s (typically [[argon]]) at the surface from an angle and sputtering material from the surface. A subclass of this is [[Focused ion beam]] milling, where [[gallium]] ions are used to produce an electron transparent membrane in a specific region of the sample, for example through a device within a microprocessor. Ion beam milling may also be used for cross-section polishing prior to SEM analysis of materials that are difficult to prepare using mechanical polishing.  
 +
* ''Conductive Coating''—An ultrathin coating of electrically-conducting material, deposited either by high vacuum evaporation or by low vacuum sputter coating of the sample. This is done to prevent the accumulation of static electric fields at the specimen due to the electron irradiation required during imaging. Such coatings include gold, gold/palladium, platinum, tungsten, graphite etc. and are especially important for the study of specimens with the scanning electron microscope. Another reason for coating, even when there is more than enough conductivity, is to improve contrast, a situation more common with the operation of a FESEM (field emission SEM). When an osmium coater is used, a layer far thinner than would be possible with any of the previously mentioned sputtered coatings is possible.<ref>SPI Supplies, [http://www.2spi.com/catalog/osmi-coat.html Osmium Plasma Coaters.] Retrieved October 5, 2008.</ref>
  
 
== Disadvantages ==
 
== Disadvantages ==
[[Image:Krillfilter2kils.jpg|thumb|Pseudocolored SEM image of the feeding basket of [[Antarctic krill]]. Real electron microscope images do not carry any color information, they are [[greyscale]]. The first degree [[filter setae]] carry in v-form two rows of second degree [[setae]], pointing towards the inside of the [[feeding basket]]. The purple ball is one micrometer in diameter. To display the total area of this structure one would have to [http://www.ecoscope.com/krill/filter/filter7/index.htm tile] 7500 times this image.]]
+
[[Image:Krillfilter2kils.jpg|thumb|Pseudocolored SEM image of the feeding basket of [[Antarctic krill]]. Real electron microscope images do not carry any color information, they are [[greyscale]]. The first degree [[filter setae]] carry in v-form two rows of second degree [[setae]], pointing towards the inside of the [[feeding basket]]. The purple ball is one micrometer in diameter. To display the total area of this structure one would have to tile this image 7500 times.]]
  
Electron microscopes are expensive to buy and maintain. They are dynamic rather than static in their operation: requiring extremely stable high-voltage supplies, extremely stable currents to each electromagnetic coil/lens, continuously-pumped high-/ultra-high-vacuum systems, and a cooling water supply circulation through the lenses and pumps. As they are very sensitive to vibration and external magnetic fields, microscopes aimed at achieving high resolutions must be housed in buildings (sometimes underground) with special services. Newer generations of TEM operating at lower voltages (around 5 kV) do not have stringent voltage supply, lens coil current, cooling water or vibration isolation requirements and as such are much less expensive to buy and far easier to install and maintain.
+
Electron microscopes are expensive to build and maintain, but the capital and running costs of confocal light microscope systems now overlaps with those of basic electron microscopes. They are dynamic rather than static in their operation, requiring extremely stable high-voltage supplies, extremely stable currents to each electromagnetic coil/lens, continuously-pumped high- or ultra-high-vacuum systems, and a cooling water supply circulation through the lenses and pumps. As they are very sensitive to vibration and external magnetic fields, microscopes designed to achieve high resolutions must be housed in stable buildings (sometimes underground) with special services such as magnetic field canceling systems. Some desktop low voltage electron microscopes have TEM capabilities at very low voltages (around 5 kV) without stringent voltage supply, lens coil current, cooling water or vibration isolation requirements and as such are much less expensive to buy and far easier to install and maintain, but do not have the same ultra-high (atomic scale) resolution capabilities as the larger instruments.
  
The samples have to be viewed in [[vacuum]], as the molecules that make up air would scatter the electrons. Recent advances have allowed hydrated samples to be imaged using an environmental scanning electron microscope
+
The samples largely have to be viewed in [[vacuum]], as the molecules that make up air would scatter the electrons. One exception is the environmental scanning electron microscope, which allows hydrated samples to be viewed in a low-pressure (up to {{convert|20|Torr|kPa|disp=s|abbr=on|lk=in}}), wet environment.
  
Scanning electron microscopes usually image conductive or semi-conductive materials best. Non-conductive materials can be imaged by an environmental scanning electron microscope. A common preparation technique is to coat the sample with a several-nanometer layer of conductive material, such as [[gold]], from a sputtering machine; however this process has the potential to disturb delicate samples.  
+
Scanning electron microscopes usually image conductive or semi-conductive materials best. Non-conductive materials can be imaged by an environmental scanning electron microscope. A common preparation technique is to coat the sample with a several-nanometer layer of conductive material, such as [[gold]], from a sputtering machine; however, this process has the potential to disturb delicate samples.  
  
The samples have to be prepared in many ways to give proper detail, which may result in ''[[Artifact (observational)|artifacts]]'' purely the result of treatment. This gives the problem of distinguishing artifacts from material, particularly in [[biology|biological]] samples. Scientists maintain that the results from various preparation techniques have been compared, and as there is no reason that they should all produce similar artifacts, it is therefore reasonable to believe that electron microscopy features correlate with living cells. In addition, higher-resolution work has been directly compared to results from [[X-ray crystallography]], providing independent confirmation of the validity of this technique. Recent work performed on unfixated, vitrified specimens has also been performed, further confirming the validity of this technique.
+
Small, stable specimens such as [[carbon nanotubes]], [[diatom]] frustules, and small mineral crystals (asbestos fibers, for example) require no special treatment before being examined in the electron microscope. Samples of hydrated materials, including almost all biological specimens have to be prepared in various ways to stabilize them, reduce their thickness (ultrathin sectioning) and increase their electron optical contrast (staining). There is a risk that these processes may result in ''[[Artifact (observational)|artifacts]],'' but these can usually be identified by comparing the results obtained by using radically different specimen preparation methods. It is generally believed by scientists working in the field that as results from various preparation techniques have been compared and that there is no reason that they should all produce similar artifacts, it is reasonable to believe that electron microscopy features correspond with those of living cells. In addition, higher-resolution work has been directly compared to results from [[X-ray crystallography]], providing independent confirmation of the validity of this technique. Since the 1980s, analysis of [[cryofixation|cryofixed]], vitrified specimens has also become increasingly used by scientists, further confirming the validity of this technique.<ref name="Adrian1984">Marc Adrian, Jacques Dubochet, Jean Lepault, and Alasdair W. McDowall, Cryo-electron microscopy of viruses, ''Nature'' 308 (5954): 32&ndash;36.</ref><ref name="Sabanay1991">I. Sabanay, T. Arad, S. Weiner, and B. Geiger, [http://jcs.biologists.org/cgi/content/abstract/100/1/227 Study of vitrified, unstained frozen tissue sections by cryoimmunoelectron microscopy,] ''Journal of Cell Science'' 100 (1): 227–236. Retrieved October 5, 2008.</ref><ref>S. Kasas, G. Dumas, G. Dietler, S. Catsicas, and M. Adrian, Vitrification of cryoelectron microscopy specimens revealed by high-speed photographic imaging, ''Journal of Microscopy'' 211 (1): 48–53.</ref>
 
 
==Electron Microscopy Application areas==
 
  
 +
==Electron microscopy application areas==
 
:{| style="background-color: transparent; width: {{{width|60%}}}"
 
:{| style="background-color: transparent; width: {{{width|60%}}}"
<p></p>
 
 
| width="{{{width|}}}" align="{{{align|left}}}" valign="{{{valign|top}}}" |
 
| width="{{{width|}}}" align="{{{align|left}}}" valign="{{{valign|top}}}" |
'''Semiconductor & Data Storage'''
+
'''Semiconductor and data storage'''
* [[Circuit Edit]]
+
* Circuit edit
* [[3D Metrology]]
+
* Defect analysis
* [[Defect Analysis]]
 
 
* [[Failure analysis]]
 
* [[Failure analysis]]
'''Biology & Life Sciences'''
+
'''Biology and life sciences'''
*[[Cryobiology]]
+
* [[Cryobiology]]
*[[Protein localization]]
+
* Protein localization
*[[Electron tomography]]
+
* [[Electron tomography]]
*[[Cellular tomography]]
+
* [[Cellular tomography]]
*[[Cryo-electron microscopy]]
+
* [[Cryo-electron microscopy]]
*[[Toxicology]]
+
* [[Toxicology]]
*[[Biological production]] and [[Viral load]] monitoring
+
* [[Biological production]] and [[viral load]] monitoring
*[[Particle analysis]]
+
* Particle analysis
*[[Pharmaceutical QC]]
+
* Pharmaceutical QC
*[[3D tissue imaging]]
+
* 3D tissue imaging
*[[Virology]]
+
* [[Virology]]
*[[Vitrification]]
+
* [[Vitrification]]
 
 
<p></p>
 
 
| width="{{{width|}}}" align="{{{align|left}}}" valign="{{{valign|top}}}" |
 
| width="{{{width|}}}" align="{{{align|left}}}" valign="{{{valign|top}}}" |
 
'''Research'''
 
'''Research'''
*[[Materials Qualification]]
+
* [[Electron beam induced deposition]]
*[[Materials and Sample Preparation]]
+
* Materials qualification
*[[Nanoprototyping]]
+
* Materials and sample preparation
*[[Nanometrology]]
+
* Nanoprototyping
*[[Device Testing and Characterization]]
+
* [[Nanometrology]]
 +
* Device testing and characterization
 
'''Industry'''
 
'''Industry'''
*[[High-Resolution Imaging]]
+
* High-resolution imaging
*[[2D & 3D Micro- Characterization]]
+
* 2D & 3D micro-characterization
*[[Macro Sample to Nanometer Metrology]]
+
* Macro sample to nanometer metrology
*[[Particle Detection and Characterization]]
+
* Particle detection and characterization
*[[Direct Beam-Writing Fabrication]]
+
* Direct beam-writing fabrication
*[[Dynamic Materials Experiments]]
+
* Dynamic materials experiments
*[[Sample Preparation]]
+
* Sample preparation
*[[Forensics]]
+
* [[Forensics]]
*[[Mining]] ([[Mineral Liberation Analysis]])
+
* [[Mining]] (mineral liberation analysis)
*[[Chemical]]/[[Petrochemical]]
+
* [[Chemical]]/[[Petrochemical]]
 
 
<p></p>
 
 
|}
 
|}
  
 
== See also ==
 
== See also ==
{{wikibookspar||The Opensource Handbook of Nanoscience and Nanotechnology}}
+
* [[Electron]]
* [[:Category:Electron microscope images]]
+
* [[Microscope]]
* [[Field emission microscope]]
+
* [[Scanning tunneling microscope]]
 +
 
 +
== Notes ==
 +
<references/>
  
 
==References==
 
==References==
<references/>
+
* Bozzola, John J., and Lonnie Dee Russell. ''Electron Microscopy.'' Sudbury, MA: Jones and Bartlett, 1998. ISBN 0763701920.
 +
* Goldstein, Joseph, et al. ''Scanning Electron Microscopy and X-ray Microanalysis.'' New York, NY: Kluwer Academic/Plenum Publishers, 2003. ISBN 0306472929.
 +
* Hunter, Elaine Evelyn. ''Practical Electron Microscopy: A Beginner's Illustrated Guide.'' New York, NY: Cambridge University Press, 1993. ISBN 0521385393.
  
 
==External links==
 
==External links==
* [http://www.2spi.com  Electron Microscopy Supplies, Accessories, and Sample Preparation Equipment] – SPI Supplies
+
All links retrieved February 13, 2024.
* [http://www.laddresearch.com Electron Microscopy Supplies] – Ladd Research
+
 
* [http://www.danilatos.com Environmental Scanning Electron Microscope (ESEM)]
+
* [http://www.danilatos.com Environmental Scanning Electron Microscope (ESEM).]  
* [http://www.microanalyst.net/index_e.phtml X-ray element analysis in electron microscope] – Information portal with X-ray microanalysis and EDX contents
+
* [http://www.microanalyst.net/index_e.phtml X-ray element analysis in electron microscope] – Information portal with X-ray microanalysis and EDX contents.
* [http://www.beemcaps.com BEEM Products for Microscopy] – BEEM, Inc.
+
* [http://www.microscopy.ethz.ch electron microscopy] Website of the ETH Zurich: Very good graphics and images, which illustrate various procedures.  
* [http://www2.physics.utoronto.ca/overview/history/microsco/microscopy.htm John H.L. Watson: Very early Electron Microscopy in the Department of Physics, the University of Toronto – A personal recollection]
+
 
* [http://americanhistory.si.edu/archives/d8452.htm Rubin Borasky Electron Microscopy Collection, 1930-1988] Archives Center, National Museum of American History, Smithsonian Institution.
+
[[Category:Physical sciences]]
 +
[[Category:Life sciences]]
  
[[ca:Microscopi electrònic]]
+
{{credit|242415873}}
[[cs:Elektronový mikroskop]]
 
[[da:Elektronmikroskop]]
 
[[de:Elektronenmikroskop]]
 
[[es:Microscopio electrónico]]
 
[[eo:Elektrona mikroskopo]]
 
[[fa:میکروسکوپ الکترونی]]
 
[[fr:Microscope électronique]]
 
[[gl:Microscopio electrónico]]
 
[[hr:Elektronski mikroskop]]
 
[[id:Mikroskop elektron]]
 
[[it:Microscopio elettronico]]
 
[[he:מיקרוסקופ אלקטרוני]]
 
[[mk:Електронски микроскоп]]
 
[[ms:Mikroskop elektron]]
 
[[nl:Elektronenmicroscopie]]
 
[[ja:電子顕微鏡]]
 
[[no:Elektronmikroskop]]
 
[[pl:Mikroskop elektronowy]]
 
[[pt:Microscópio eletrônico]]
 
[[ru:Электронный микроскоп]]
 
[[simple:Electron microscope]]
 
[[fi:Elektronimikroskooppi]]
 
[[sv:Elektronmikroskop]]
 
[[ta:இலத்திரன் நுண்நோக்கி]]
 
[[te:ఎలక్ట్రాన్ సూక్ష్మదర్శిని]]
 
[[vi:Kính hiển vi điện tử]]
 
[[zh:電子顯微鏡]]
 
{{credit|170215286}}
 

Latest revision as of 15:58, 13 February 2024

An electron microscope.

An electron microscope is a type of microscope that uses electrons to illuminate a specimen and create an enlarged image. Electron microscopes have much greater resolving power than light microscopes and can obtain much higher magnifications. Some electron microscopes can magnify specimens up to 2 million times, while the best light microscopes are limited to magnifications of 2000 times. Both electron and light microscopes have resolution limitations, imposed by their wavelength. The greater resolution and magnification of the electron microscope is due to the wavelength of an electron, its de Broglie wavelength, being much smaller than that of a light photon, electromagnetic radiation.

The electron microscope uses electrostatic and electromagnetic lenses in forming the image by controlling the electron beam to focus it at a specific plane relative to the specimen in a manner similar to how a light microscope uses glass lenses to focus light on or through a specimen to form an image.

History

Electron microscope constructed by Ernst Ruska in 1933.
Did you know?
The first electron microscope prototype was built in 1931 by German engineers Ernst Ruska and Max Knoll

The first electron microscope prototype was built in 1931 by German engineers Ernst Ruska and Max Knoll.[1] Although this initial instrument was only capable of magnifying objects by four hundred times, it demonstrated the principles of an electron microscope. Two years later, Ruska constructed an electron microscope that exceeded the resolution possible using an optical microscope.[1]

Reinhold Rudenberg, the research director of Siemens, had patented the electron microscope in 1931, although Siemens was doing no research on electron microscopes at that time. In 1937, Siemens began funding Ruska and Bodo von Borries to develop an electron microscope. Siemens also employed Ruska's brother Helmut to work on applications, particularly with biological specimens.[1][2]

In the same decade Manfred von Ardenne pioneered the scanning electron microscope and his universal electron microscope.[3]

Siemens produced the first commercial TEM in 1939, but the first practical electron microscope had been built at the University of Toronto in 1938, by Eli Franklin Burton and students Cecil Hall, James Hillier, and Albert Prebus.[4]

Although modern electron microscopes can magnify objects up to two million times, they are still based upon Ruska's prototype. The electron microscope is an integral part of many laboratories. Researchers use them to examine biological materials (such as microorganisms and cells), a variety of large molecules, medical biopsy samples, metals and crystalline structures, and the characteristics of various surfaces. The electron microscope is also used extensively for inspection, quality assurance and failure analysis applications in industry, including, in particular, semiconductor device fabrication.

Types

Transmission Electron Microscope (TEM)

Diagram of a transmission electron microscope.

The original form of electron microscopy, Transmission electron microscopy (TEM) involves a high voltage electron beam emitted by an electron gun, usually fitted with a tungsten filament cathode as the electron source. The electron beam is accelerated by an anode typically at +100keV (40 to 400 keV) with respect to the cathode, focused by electrostatic and electromagnetic lenses, and transmitted through a specimen that is in part transparent to electrons and in part scatters them out of the beam. When it emerges from the specimen, the electron beam carries information about the structure of the specimen that is magnified by the objective lens system of the microscope. The spatial variation in this information (the "image") is recorded by projecting the magnified electron image onto a fluorescent viewing screen coated with a phosphor or scintillator material such as zinc sulfide. The image can be photographically recorded by exposing a photographic film or plate directly to the electron beam, or a high-resolution phosphor may be coupled by means of a fiber optic light-guide to the sensor of a CCD (charge-coupled device) camera. The image detected by the CCD may be displayed on a monitor or computer.

Resolution of the TEM is limited primarily by spherical aberration, but a new generation of aberration correctors have been able to partially overcome spherical aberration to increase resolution. Software correction of spherical aberration for the High Resolution TEM HRTEM has allowed the production of images with sufficient resolution to show carbon atoms in diamond separated by only 0.89 ångström (89 picometers) and atoms in silicon at 0.78 ångström (78 picometers)[5][6] at magnifications of 50 million times.[7] The ability to determine the positions of atoms within materials has made the HRTEM an important tool for nano-technologies research and development.

Scanning Electron Microscope (SEM)

Unlike the TEM, where electrons of the high voltage beam form the image of the specimen, the Scanning Electron Microscope (SEM)[8] produces images by detecting low energy secondary electrons which are emitted from the surface of the specimen due to excitation by the primary electron beam. In the SEM, the electron beam is rastered across the sample, with detectors building up an image by mapping the detected signals with beam position.

An image of an ant from a scanning electron microscope

Generally, the TEM resolution is about an order of magnitude greater than the SEM resolution, however, because the SEM image relies on surface processes rather than transmission it is able to image bulk samples and has a much greater depth of view, and so can produce images that are a good representation of the 3D structure of the sample.

Reflection Electron Microscope (REM)

In the Reflection Electron Microscope (REM) as in the TEM, an electron beam is incident on a surface, but instead of using the transmission (TEM) or secondary electrons (SEM), the reflected beam of elastically scattered electrons is detected. This technique is typically coupled with Reflection High Energy Electron Diffraction and Reflection high-energy loss spectrum (RHELS). Another variation is Spin-Polarized Low-Energy Electron Microscopy (SPLEEM), which is used for looking at the microstructure of magnetic domains.[9]

Scanning Transmission Electron Microscope (STEM)

The STEM rasters a focused incident probe across a specimen that (as with the TEM) has been thinned to facilitate detection of electrons scattered through the specimen. The high resolution of the TEM is thus possible in STEM. The focusing action (and aberrations) occur before the electrons hit the specimen in the STEM, but afterward in the TEM. The STEM's use of SEM-like beam rastering simplifies annular dark-field imaging, and other analytical techniques, but also means that image data is acquired in serial rather than in parallel fashion.

Sample preparation

An insect coated in gold for viewing with a scanning electron microscope.

Materials to be viewed under an electron microscope may require processing to produce a suitable sample. The technique required varies depending on the specimen and the analysis required:

  • Chemical Fixation for biological specimens aims to stabilize the specimen's mobile macromolecular structure by chemical crosslinking of proteins with aldehydes such as formaldehyde and glutaraldehyde, and lipids with osmium tetroxide.
  • Cryofixation—freezing a specimen so rapidly, to liquid nitrogen or even liquid helium temperatures, that the water forms vitreous (non-crystalline) ice. This preserves the specimen in a snapshot of its solution state. An entire field called cryo-electron microscopy has branched from this technique. With the development of cryo-electron microscopy of vitreous sections (CEMOVIS), it is now possible to observe virtually any biological specimen close to its native state.
  • Dehydrationfreeze drying, or replacement of water with organic solvents such as ethanol or acetone, followed by critical point drying or infiltration with embedding resins.
  • Embedding, biological specimens—infiltration of the tissue with a resin such as Araldite epoxy or acrylic resin followed by ultra-thin sectioning and staining.
  • Embedding, materials—After embedding in resin, the specimen is usually ground and polished to a mirror-like finish using ultra-fine abrasives. The polishing process must be performed carefully to minimize scratches and other polishing artifacts that reduce image quality.
  • Sectioning—produces thin slices of specimen, semitransparent to electrons. These can be cut on an ultramicrotome with a diamond knife to produce ultrathin slices about 90 nm thick. Glass knives are also used because they can be made in the lab and are much cheaper.
  • Staining—uses heavy metals such as lead, uranium or tungsten to scatter imaging electrons and thus give contrast between different structures, since many (especially biological) materials are nearly "transparent" to electrons (weak phase objects). In biology, specimens are usually stained "en bloc" before embedding and also later stained directly after sectioning by brief exposure to aqueous (or alcoholic) solutions of the heavy metal stains.
  • Freeze-fracture or freeze-etch—a preparation method particularly useful for examining lipid membranes and their incorporated proteins in "face on" view. The fresh tissue or cell suspension is frozen rapidly (cryofixed), then fractured by simply breaking or by using a microtome while maintained at liquid nitrogen temperature. The cold fractured surface (sometimes "etched" by increasing the temperature to about -100°C for several minutes to let some ice sublime) is then shadowed with evaporated platinum or gold at an average angle of 45° in a high vacuum evaporator. A second coat of carbon, evaporated perpendicular to the average surface plane is often performed to improve stability of the replica coating. The specimen is returned to room temperature and pressure, then the extremely fragile "pre-shadowed" metal replica of the fracture surface is released from the underlying biological material by careful chemical digestion with acids, hypochlorite solution or SDS detergent. The still-floating replica is thoroughly washed from residual chemicals, carefully fished up on EM grids, dried then viewed in the TEM.
  • Ion Beam Milling—thins samples until they are transparent to electrons by firing ions (typically argon) at the surface from an angle and sputtering material from the surface. A subclass of this is Focused ion beam milling, where gallium ions are used to produce an electron transparent membrane in a specific region of the sample, for example through a device within a microprocessor. Ion beam milling may also be used for cross-section polishing prior to SEM analysis of materials that are difficult to prepare using mechanical polishing.
  • Conductive Coating—An ultrathin coating of electrically-conducting material, deposited either by high vacuum evaporation or by low vacuum sputter coating of the sample. This is done to prevent the accumulation of static electric fields at the specimen due to the electron irradiation required during imaging. Such coatings include gold, gold/palladium, platinum, tungsten, graphite etc. and are especially important for the study of specimens with the scanning electron microscope. Another reason for coating, even when there is more than enough conductivity, is to improve contrast, a situation more common with the operation of a FESEM (field emission SEM). When an osmium coater is used, a layer far thinner than would be possible with any of the previously mentioned sputtered coatings is possible.[10]

Disadvantages

Pseudocolored SEM image of the feeding basket of Antarctic krill. Real electron microscope images do not carry any color information, they are greyscale. The first degree filter setae carry in v-form two rows of second degree setae, pointing towards the inside of the feeding basket. The purple ball is one micrometer in diameter. To display the total area of this structure one would have to tile this image 7500 times.

Electron microscopes are expensive to build and maintain, but the capital and running costs of confocal light microscope systems now overlaps with those of basic electron microscopes. They are dynamic rather than static in their operation, requiring extremely stable high-voltage supplies, extremely stable currents to each electromagnetic coil/lens, continuously-pumped high- or ultra-high-vacuum systems, and a cooling water supply circulation through the lenses and pumps. As they are very sensitive to vibration and external magnetic fields, microscopes designed to achieve high resolutions must be housed in stable buildings (sometimes underground) with special services such as magnetic field canceling systems. Some desktop low voltage electron microscopes have TEM capabilities at very low voltages (around 5 kV) without stringent voltage supply, lens coil current, cooling water or vibration isolation requirements and as such are much less expensive to buy and far easier to install and maintain, but do not have the same ultra-high (atomic scale) resolution capabilities as the larger instruments.

The samples largely have to be viewed in vacuum, as the molecules that make up air would scatter the electrons. One exception is the environmental scanning electron microscope, which allows hydrated samples to be viewed in a low-pressure (up to 20 Torr/2.7 kPa), wet environment.

Scanning electron microscopes usually image conductive or semi-conductive materials best. Non-conductive materials can be imaged by an environmental scanning electron microscope. A common preparation technique is to coat the sample with a several-nanometer layer of conductive material, such as gold, from a sputtering machine; however, this process has the potential to disturb delicate samples.

Small, stable specimens such as carbon nanotubes, diatom frustules, and small mineral crystals (asbestos fibers, for example) require no special treatment before being examined in the electron microscope. Samples of hydrated materials, including almost all biological specimens have to be prepared in various ways to stabilize them, reduce their thickness (ultrathin sectioning) and increase their electron optical contrast (staining). There is a risk that these processes may result in artifacts, but these can usually be identified by comparing the results obtained by using radically different specimen preparation methods. It is generally believed by scientists working in the field that as results from various preparation techniques have been compared and that there is no reason that they should all produce similar artifacts, it is reasonable to believe that electron microscopy features correspond with those of living cells. In addition, higher-resolution work has been directly compared to results from X-ray crystallography, providing independent confirmation of the validity of this technique. Since the 1980s, analysis of cryofixed, vitrified specimens has also become increasingly used by scientists, further confirming the validity of this technique.[11][12][13]

Electron microscopy application areas

Semiconductor and data storage

  • Circuit edit
  • Defect analysis
  • Failure analysis

Biology and life sciences

  • Cryobiology
  • Protein localization
  • Electron tomography
  • Cellular tomography
  • Cryo-electron microscopy
  • Toxicology
  • Biological production and viral load monitoring
  • Particle analysis
  • Pharmaceutical QC
  • 3D tissue imaging
  • Virology
  • Vitrification

Research

  • Electron beam induced deposition
  • Materials qualification
  • Materials and sample preparation
  • Nanoprototyping
  • Nanometrology
  • Device testing and characterization

Industry

  • High-resolution imaging
  • 2D & 3D micro-characterization
  • Macro sample to nanometer metrology
  • Particle detection and characterization
  • Direct beam-writing fabrication
  • Dynamic materials experiments
  • Sample preparation
  • Forensics
  • Mining (mineral liberation analysis)
  • Chemical/Petrochemical

See also

Notes

  1. 1.0 1.1 1.2 Ernst Ruska, Ernst Ruska Nobel Prize autobiography, Nobel Prize. Retrieved October 5, 2008.
  2. D.H. Kruger, P. Schneck, and H.R. Gelderblom, Helmut Ruska and the visualization of viruses, The Lancet 355 (9216): 1713–1717.
  3. M. von Ardenne and D. Beischer, Untersuchung von metalloxud-rauchen mit dem universal-elektronenmikroskop, Zeitschrift Electrochemie 46: 270–277.
  4. Lemelson-MIT Program, MIT biography of Hillier.
  5. Paul Preuss, OÅM: World-Record Resolution at 0.78 Å, Berkeley Lab Currents.
  6. P.D. Nellist, M.F. Chisholm, N. Dellby, O.L. Krivanek, M.F. Murfitt, Z.S. Szilagyi, A.R. Lupini, A. Borisevich, W.H. Sides, Jr., and S.J. Pennycook, Direct Sub-Angstrom Imaging of a Crystal Lattice, Science 305 (5691): 1741. Retrieved October 5, 2008.
  7. Department of Energy, The Scale of Things, Office of Basic Energy Sciences (BES). Retrieved October 5, 2008.
  8. D. McMullan, Scanning Electron Microscopy, 1928-1965, Microscopy Society of America. Retrieved October 5, 2008.
  9. National Center for Electron Microscopy, SPLEEM. Retrieved October 5, 2008.
  10. SPI Supplies, Osmium Plasma Coaters. Retrieved October 5, 2008.
  11. Marc Adrian, Jacques Dubochet, Jean Lepault, and Alasdair W. McDowall, Cryo-electron microscopy of viruses, Nature 308 (5954): 32–36.
  12. I. Sabanay, T. Arad, S. Weiner, and B. Geiger, Study of vitrified, unstained frozen tissue sections by cryoimmunoelectron microscopy, Journal of Cell Science 100 (1): 227–236. Retrieved October 5, 2008.
  13. S. Kasas, G. Dumas, G. Dietler, S. Catsicas, and M. Adrian, Vitrification of cryoelectron microscopy specimens revealed by high-speed photographic imaging, Journal of Microscopy 211 (1): 48–53.

References
ISBN links support NWE through referral fees

  • Bozzola, John J., and Lonnie Dee Russell. Electron Microscopy. Sudbury, MA: Jones and Bartlett, 1998. ISBN 0763701920.
  • Goldstein, Joseph, et al. Scanning Electron Microscopy and X-ray Microanalysis. New York, NY: Kluwer Academic/Plenum Publishers, 2003. ISBN 0306472929.
  • Hunter, Elaine Evelyn. Practical Electron Microscopy: A Beginner's Illustrated Guide. New York, NY: Cambridge University Press, 1993. ISBN 0521385393.

External links

All links retrieved February 13, 2024.

Credits

New World Encyclopedia writers and editors rewrote and completed the Wikipedia article in accordance with New World Encyclopedia standards. This article abides by terms of the Creative Commons CC-by-sa 3.0 License (CC-by-sa), which may be used and disseminated with proper attribution. Credit is due under the terms of this license that can reference both the New World Encyclopedia contributors and the selfless volunteer contributors of the Wikimedia Foundation. To cite this article click here for a list of acceptable citing formats.The history of earlier contributions by wikipedians is accessible to researchers here:

The history of this article since it was imported to New World Encyclopedia:

Note: Some restrictions may apply to use of individual images which are separately licensed.